The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): A phantom study

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [a]Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, No. 45 Changchun Street, Beijing 100053, PR China [b]Siemens Ltd., China, No. 278, Zhouzhu Road, Pudong New Area, Shanghai 201318, PR China
出处:
ISSN:

关键词: C-arm flat panel CT Multi-slice CT Radiation dose

摘要:
Purpose: To assess and compare the radiation dose of DynaCT in phantoms to that of MSCT. Methods and materials: A male Anderson Radiation Therapy 200 phantom with embedded thermoluminescence dosimeters (TLDs) was scanned with DynaCT (Axiom Artis dTA, Siemens Healthcare, Forchheim, Germany) and MSCTs (MSCT1: Lightspeed VCT, GE, Milwaukee, USA; MSCT2: Sensation Cardiac 64, Siemens Healthcare, Erlangen, Germany), respectively. For DynaCT, radiation exposure data with automatic exposure control were obtained from scanning of head, chest and abdomen. For MSCT, scanning protocols for head, chest, and abdomen were examined under conventional exposure conditions. Effective doses were calculated from LiF-TLD measurements according to ICRP103. A Catphan phantom (Phantom Lab, USA) was used to evaluate the spatial resolution and low contrast detectability of DynaCT and MSCT. Results: The effective doses for DynaCT from head, chest and abdomen scanning were 1.18, 7.32, 7.48 mSv (20 s scan) and 0.85, 6.01, 7.04 mSv (8 s scan); the effective doses for MSCT from head, chest and abdomen scanning were 3.33, 7.62 and 8.42 mSv (MSCT1), and 1.89, 7.52, 8.23 mSv (MSCT2). Significant difference between the organ doses from DynaCT and from MSCT (p<.05) was shown. The spatial resolution of 12 lp/cm was achieved and it was able to recognize a 3 mm low contrast object at 0.5% contrast level in DynaCT, which was on the same level as in the MSCT images. Conclusion: The phantom study indicates that DynaCT applies significantly less dose to patient and achieves similar spatial resolution and low contrast detectability to standard diagnostic MSCT. (C) 2011 Elsevier Ireland Ltd. All rights reserved.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2011]版:
大类 | 3 区 医学
小类 | 3 区 核医学
最新[2023]版:
大类 | 3 区 医学
小类 | 3 区 核医学
JCR分区:
出版当年[2010]版:
Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2023]版:
Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING

影响因子: 最新[2023版] 最新五年平均 出版当年[2010版] 出版当年五年平均 出版前一年[2009版] 出版后一年[2011版]

第一作者:
第一作者机构: [a]Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, No. 45 Changchun Street, Beijing 100053, PR China
通讯作者:
通讯机构: [a]Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, No. 45 Changchun Street, Beijing 100053, PR China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:16409 今日访问量:0 总访问量:869 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学宣武医院 技术支持:重庆聚合科技有限公司 地址:北京市西城区长椿街45号宣武医院