Finite Element Analysis of Influence of Axial Position of Center of Rotation of a Cervical Total Disc Replacement on Biomechanical Parameters: Simulated 2-Level Replacement Based on a Validated Model.

文献详情

资源类型:
机构: [1]Department of Mechanical Engineering, Tsinghua University, Beijing, China. [2]Biomechanics and Biotechnology Lab, Research Institute of Tsinghua University in Shenzhen, Shenzhen, Guangdong, China. [3]National Research Center for Rehabilitation Technical Aids, Beijing, China. [4]Department of Mechanical Engineering, Tsinghua University, Beijing, China; Biomechanics and Biotechnology Lab, Research Institute of Tsinghua University in Shenzhen, Shenzhen, Guangdong, China. Electronic address: weiqliu@hotmail.com.
出处:
ISSN:

摘要:
Finite element models have been widely used to predict biomechanical parameters of the cervical spine. Previous studies investigated the influence of position of rotational centers of prostheses on cervical biomechanical parameters after 1-level total disc replacement. The purpose of this study was to explore the effects of axial position of rotational centers of prostheses on cervical biomechanics after 2-level total disc replacement. A validated finite element model of C3-C7 segments and 2 prostheses, including the rotational center located at the superior endplate (SE) and inferior endplate (IE), was developed. Four total disc replacement models were used: 1) IE inserted at C4-C5 disc space and IE inserted at C5-C6 disc space (IE-IE), 2) IE-SE, 3) SE-IE, and 4) SE-SE. All models were subjected to displacement control combined with a 50 N follower load to simulate flexion and extension motions in the sagittal plane. For each case, biomechanical parameters, including predicted moments, range of rotation at each level, facet joint stress, and von Mises stress on the ultra-high-molecular-weight polyethylene core of the prostheses, were calculated. The SE-IE model resulted in significantly lower stress at the cartilage level during extension and at the ultra-high-molecular-weight polyethylene cores when compared with the SE-SE construct and did not generate hypermotion at the C4-C5 level compared with the IE-SE and IE-IE constructs. Based on the present analysis, the SE-IE construct is recommended for treating cervical disease at the C4-C6 level. This study may provide a useful model to inform clinical operations. Copyright © 2017 Elsevier Inc. All rights reserved.

语种:
PubmedID:
中科院(CAS)分区:
出版当年[2016]版:
大类 | 3 区 医学
小类 | 3 区 临床神经病学 3 区 外科
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 临床神经病学 4 区 外科
第一作者:
第一作者机构: [1]Department of Mechanical Engineering, Tsinghua University, Beijing, China.
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:17292 今日访问量:0 总访问量:929 更新日期:2025-06-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学宣武医院 技术支持:重庆聚合科技有限公司 地址:北京市西城区长椿街45号宣武医院