机构:[1]Department of Neurosurgery, Beijing Hospital, Beijing 100730, China[2]Department of Neurosurgery, Xuan Wu Hospital, Beijing 100053, China神经外科首都医科大学宣武医院[3]Department of Neurosurgery, Second Hospital of Dalian, Dalian 116011, China
Objective: To explore the causes, prevention, and management of the complications during intracranial aneurysm embolization with controllable coils (mechanical detachable spiral, MDS; and Guglielmi detachable coil, GDC). Methods: Retrospective review of 120 cases with 125 intracranial aneurysms embolized with controllable coils from March 1995 to July 1999 was conducted. The 20 accidents (in 18 cases) including aneurysm rupture, over-embolization, protrusion of coil end into the parent artery, and thrombosis of the parent artery were analyzed. Results: Among the 20 accidents, there were 6 aneurysm ruptures, 6 over-embolizations (in 5 cases), 6 coil protrusions, and 2 thromboses (one was secondary to coil protrusion). The embolization-related mortality was 3.33% (4/120), the permanent neurological deficit was 1.67% (2/120), and the transitory neurological deficit was 3.33% (4/120). The occurrence and outcome of the complications were related to the embolizing technique, the pattern of aneurysm and its parent artery, the imperfection of embolic materials, and the observation and management during embolization. Conclusion: Skilled embolizing technique, better understanding of the angio-anatomy of an aneurysm and its parent artery, correct judgement and management during embolization, and improvement of embolic materials are beneficial to the reduction of complications and to the melioration of the outcome of complications.
语种:
外文
第一作者:
第一作者机构:[1]Department of Neurosurgery, Beijing Hospital, Beijing 100730, China
通讯作者:
通讯机构:[1]Department of Neurosurgery, Beijing Hospital, Beijing 100730, China
推荐引用方式(GB/T 7714):
王大明,凌锋,王安顺.Complication analysis of intracranial aneurysm embolization with controllable coils[J].Chinese Medical Sciences Journal.2004,19(1):
APA:
王大明,凌锋&王安顺.(2004).Complication analysis of intracranial aneurysm embolization with controllable coils.Chinese Medical Sciences Journal,19,(1)
MLA:
王大明,et al."Complication analysis of intracranial aneurysm embolization with controllable coils".Chinese Medical Sciences Journal 19..1(2004)