当前位置: 首页 > 详情页

Effect of echo time on the quantification of diffusion tensor imaging of brain at 3.0T

| 认领 | 导出 |

文献详情

资源类型:

收录情况: ◇ 统计源期刊 ◇ 北大核心 ◇ CSCD-C

机构: [1]首都医科大学宣武医院放射科,北京100053
出处:
ISSN:

关键词: Diffusion tensor imaging Echo time Quantitative study Signal to noise ratio

摘要:
Objective: To investigate the effects of echo time (TE) on diffusion tensor quantification of brain. Methods: Seven Macaca mulatta underwent diffusion tensor MR imaging (DTI) with a series of TEs at 3.0T MR scanners. The diffusion scalars including mean diffusivity (MD), fractional anisotropy (FA), and eigenvalues (λ 1, λ 2 and λ 3) were measured in the region of interest (ROI) at the bilateral posterior limb of internal capsule and the putamen. Pearson correlation coefficients were used to test the relationships between TE and the diffusion scalars. Results: The MD showed positive correlation with TE in gray matter (GM) (P<0.001) but not in white matter (WM) (P>0.05). Both the FA and λ 1 of GM and WM were positively correlated with TE (P<0.001). The λ 2 of GM increased and λ 2 of WM decreased with increasing TE. And the λ 3 were negatively correlated with TE. Conclusion: TE may influence the quantification of diffusion tensor imaging, which should be controlled in multi-center studies.

语种:
第一作者:
第一作者机构: [1]首都医科大学宣武医院放射科,北京100053
通讯作者:
通讯机构: [1]首都医科大学宣武医院放射科,北京100053
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:16409 今日访问量:0 总访问量:869 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学宣武医院 技术支持:重庆聚合科技有限公司 地址:北京市西城区长椿街45号宣武医院