Diagnostic performance of 320-detector CT coronary angiography in patients with atrial fibrillation: preliminary results

文献详情

资源类型:

收录情况: ◇ SCIE

机构: [1]Capital Med Univ, Dept Radiol, Beijing Anzhen Hosp, Beijing 100029, Peoples R China; [2]Capital Med Univ, Dept Radiol, Beijing Anzhen Hosp, 2 Anzhen Rd, Beijing 100029, Peoples R China
出处:
ISSN:

关键词: 320-MDCT Coronary arteries Coronary artery disease Cardiac CT Atrial fibrillation

摘要:
To evaluate the feasibility, diagnostic accuracy, and radiation dose of CT coronary angiography (CTCA) in patients with atrial fibrillation (AF) using 320-detector CT. Thirty-seven patients with persistent AF and suspected coronary artery disease (CAD) were enrolled. All patients underwent both 320-detector CTCA and conventional coronary angiography (CCA). CT image quality and the presence of significant (>= 50%) stenosis were evaluated by two radiologists blinded to the results of CCA. The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were calculated using CCA as the reference standard. Differences in detection of coronary artery stenosis between 320-detector CTCA and CCA were evaluated with McNemar's test. Patient radiation dose was calculated by multiplying dose length product by conversion coefficient of 0.017. In total 474 evaluated coronary segments, 459 (96.8%) segments were diagnostically evaluable. On per-segment analysis, sensitivity, specificity, PPV and NPV were 90.0% (18 of 20), 99.3% (436 of 439), 85.7% (18 of 21) and 99.5% (436 of 438). No significant difference was found between 320-detector CTCA and CCA on the detection of significant stenosis (P = 1.000). Effective doses of 320-detector CTCA was 13.0 +/- 4.7 mSv. 320-detector CTCA is feasible and accurate in excluding CAD in patients with AF.

语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2010]版:
大类 | 2 区 医学
小类 | 2 区 核医学
最新[2023]版:
大类 | 2 区 医学
小类 | 2 区 核医学
JCR分区:
出版当年[2009]版:
Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2023]版:
Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING

影响因子: 最新[2023版] 最新五年平均 出版当年[2009版] 出版当年五年平均 出版前一年[2008版] 出版后一年[2010版]

第一作者:
第一作者机构: [1]Capital Med Univ, Dept Radiol, Beijing Anzhen Hosp, Beijing 100029, Peoples R China;
通讯作者:
通讯机构: [1]Capital Med Univ, Dept Radiol, Beijing Anzhen Hosp, Beijing 100029, Peoples R China; [2]Capital Med Univ, Dept Radiol, Beijing Anzhen Hosp, 2 Anzhen Rd, Beijing 100029, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:16409 今日访问量:0 总访问量:869 更新日期:2025-01-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 首都医科大学宣武医院 技术支持:重庆聚合科技有限公司 地址:北京市西城区长椿街45号宣武医院